Features
and specifications
- Capacity: 32,530 measurements total
- User-selectable sampling intervals: 0.5 seconds
to 9 hours, recording times up to 1 year
- Readout and relaunch with optional HOBO
Shuttle
- Drop-proof to 5'
- Mounting kit included (hook/loop, magnet, and
tape)
- Programmable start time/date
- Memory modes: stop when full, wrap-around when
full
- Nonvolatile EEPROM memory retains data even if
battery fails
- Blinking LED confirms operation
- User-replaceable battery lasts 1 year
- Battery level indication at launch
- Operating range: 0°C to +50°C (+32°F to +122°F),
0 to 95% relative humidity, non-condensing, non-fogging
- Time accuracy: ±1 minute per week at +20°C (+68°F)
- Size/Weight: 2.4 x 1.9 x 0.8" (68 x 48 x 19
mm)/ approx. 0.9 oz. (26 gms)
Internal Temperature Sensor
This sensor is also used to measure the cold-junction temperature. The
thermocouple readings are compensated in the logger regardless of whether
this channel is selected.
- Internal Range: 0°C to +50°C (+32°F to +122°F)
- Internal Resolution: 0.4°C @ +21°C (0.7°F @
+70°F)
- Internal Accuracy: ±0.7°C @ +21°C (±1.3°F @
+70°F)

Thermocouple Probes and Wires
All probes have standard subminiature thermocouple connectors
and plug directly into HOBO Thermocouple Family loggers. The connectors
are color-coded for type J (black), K (yellow), and T (blue) thermocouples
according to the ANSI standard. Note that the accuracy specifications are
for the probes only, and must be added to the logger's accuracy to
determine the combined measurement accuracy.
Beaded Thermocouple Probes on Spool Caddies
These exposed-junction thermocouple probes include 6' of insulated 30-AWG
wire wound on an integrated spool caddy/subminiature connector.
Type
J
Range: +32°F to +482°F (0°C to +250°C)
Initial Accuracy: ±2°F (±1.1°C)
Teflon Insulation*
Type K
Range:
+32°F to +545°F (0°C to +285°C)
Initial Accuracy: ±2°F (±1.1°C)
Range: +545°F to +900°F (+285°C to +482°C)
Initial Accuracy: ±0.4%***
Glass Braid Insulation**
Type
T
Range: -330°F to -85°F
(-200°C to -65°C)
Initial Accuracy: ±0.8%***
Range: -85°F to +270°F (-65°C to +130°C)
Initial Accuracy: ±0.9°F (±0.5°C)
Range: +270°F to +356°F (+130°C to +180°C)
Initial Accuracy: ±0.4%***
Teflon Insulation*
High-Temperature Stainless-Steel Probe - Type K
. 12" 304 stainless-steel probe; 3/16" diameter with isolated
thermocouple
. 1" handle with spring-type strain relief
. 6' 19-AWG cable
Range: +32°F to +559°F (0°C to +293°C)
Initial Accuracy: ±4°F (±2.2°C)
Range: +559°F to +1650°F (+293°C to +900°C)****
Initial Accuracy: ±0.75%***
Subminiature Thermocouple Connectors
. Accept stranded or solid wire up to 20 AWG
. Internal silicone grommet for strain relief
. Captive cover screws
. Write-on window
. Type J, K, and T versions available
|
Measurement
Specifications
(Each model has two software-selectable ranges)
HOBO Type J Thermocouple (part number H12-001)
| Range |
Max./Min. Res. |
Accuracy |
0°C to +250°C
(+32°F to +482°F) |
1°C / 2°C
(1.8°F / 3.6°F) |
±3.2°C @ +125°C
(±5.7°F @ +257°F) |
0°C to +750°C
(+32°F to +1382°F) |
3°C / 6°C
(5.4°F / 11°F) |
±8.5°C @ +375°C
(±15.3°F @ +707°F) |

HOBO Type K Thermocouple (part number H12-002)
| Range |
Max./Min. Res. |
Accuracy |
0°C to +500°C
(+32°F to +932°F) |
2°C / 4°C
(3.6°F / 7°F) |
±6°C @ +250°C
(±10.8°F @ +482°F) |
0°C to +1250°C
(+32°F to +2282°F) |
5°C / 10°C
(9°F / 18°F) |
±14°C @ +625°C
(±25°F @ +1157°F) |

HOBO Type T Thermocouple (part number H12-003)
| Range |
Max./Min. Res. |
Accuracy |
-200°C to +50°C
(-328°F to +122°F) |
1°C / 6°C
(1.8°F / 11°F) |
±4°C @ -75°C
(±7.2°F @ -103°F) |
-70°C to +180°C
(-94°F to +356°F) |
1°C / 3°C
(1.8°F / 5.4°F) |
±3°C @ +55°C
(±5.4°F @ +131°F) |

Note: The accuracy plots show the worst
case error for the different loggers and their ranges. Please note that
this does not include any error associated with the attached thermocouple.
Thermocouple Probes and Wires
(continued)
Thermocouple Wires
100' coils of insulated 24-AWG thermocouple wire

Type J
Range: +32°F to +482°F (0°C to +250°C)
Initial Accuracy: ±4°F (±2.2°C)
Teflon Insulation*
Type K
Range: +32°F to +545°F (0°C to +285°C)
Initial Accuracy: ±4°F (±2.2°C)
Range: +545°F to +900°F (+285°C to +482°C)
Initial Accuracy: ±0.75%***
Glass Braid Insulation**
Type T
Range: -330°F to -85°F (-200°C to -65°C)
Initial Accuracy: ±1.5%***
Range: -85°F to +270°F (-65°C to +130°C)
Initial Accuracy: ±1.8°F (±1°C)
Range: +270°F to +356°F (+130°C to +180°C)
Initial Accuracy: ±0.75%***
Teflon Insulation*
Warranty: Thermocouple Loggers are
intended for indoor use. They have not been tested or are not guaranteed
for outdoor use. Thermocouple probes, connectors, and wires are warranted
for manufacturing defects on receipt only.
* Teflon insulation is rated up to +400°F (+200°C) continuous, and +482°F
(+250°C) for 24 hours.
** Glass braid is rated up to +900°F (+482°C) continuous, and +1000°F
(+537°C) for single exposure.
*** Percentage is of degrees above or below ice point (+32°F or 0°C)
**** Probe is rated up to +1650°F (+900°C) continuous, and +2282°F
(+1235°C) for one-hour exposures.
|